An assembly of 24 double pancake coils has been used to demonstrate experimentally the possibility of creating a homogeneous field of MRI quality using double pancakes and to test the assembly procedures planned for Iseult. The field of this 1.5 T magnet has been measured at 648 separate points using an NMR probe. The measured homogeneity of 300 ppm peak-to-peak is in line with theoretical predictions.
• superconducting magnet physics and technology › Superconducting Magnets for Particle Physics Research
• Accelerators, Cryogenics and Magnetism Division (DACM)
• LCSE • pas de titre